LR-Cal Kalibriergeräte / Calibration Equipment von / from DRUCK & TEMPERATUR Leitenberger - GERMANY

Pressure Controller LR-Cal LCC 200Electronic Pressure Controller and Calibrator for small Pressure, Vacuum and Differential pressure

Model LR-Cal LCC 200

  • Pneumatic, measures pressure, vacuum and differential pressure
  • Generates Pressure/Vacuum -600 mbar to +1000 mbar (built-in electrical pump)
  • Pressure ranges to 1 mbar, 10 mbar, 100 mbar and 1000 mbar
  • Accuracy up to ±0.05% FS (+1 digit)
  • Great adjustment accuracy 0.02% FS
  • Chargeable Li-Ion battery
  • Generates pressure/vacuum by pressing a key
  • Touch screen operation

Important Note:
This instrument is not available in the moment.

Description

Description LR-Cal LCC 200

The LR-Cal LCC 200 pressure controller measures and sources pressure and vacuum and can be used as reference pressure instrument as well as calibrator for pressure reading instruments like presure gauges, pressure transmitters, pressure switches, etc. Thanks to internal pressure/vacuum generation, no auxiliary tools are required for operation. An internal Li-ion battery (chargeable) makes mobile operation very easy. Parallel operation of mains and battery supply allows for maximum flexibility.

Typical Applications:

  • Mobile and stationary pressure standard for low pressure and vacuum
  • Mobile and stationary of all kind of pressure reading instruments
  • Mobile and stationary pressure and vacuum source for calibration purposes
  • Leak test

Standard scope of delivery:

  • Pressure Controller LR-Cal LCC 200 with built-in Li-Ion battery
  • Mains adapter 110...230 VAC
  • 1 piece of transparent pneumatic hose/tube
  • Operating manual

Options: see "Accessories"

Back view:
Back side of LR-Cal LCC 200 pressure controller

Screenshot main screen:
Screenshot LR-Cal LCC 200 pressure controller

Specification

Specification LR-Cal LCC 200

Type LCC 200-1 LCC 200-10 LCC 200-100 LCC 200-1000
Order-Code LCC-200-1 LCC-200-10 LCC-200-100 LCC-200-1000
Pressure range -1...+1 mbar -10...+10 mbar -100...+100 mbar -1000...+1000 mbar
Overpressure For delta p measurements max. system pressure 2 bar
Measurement Uncertainty (+1 digit) ±0.15% FS ±0.05% FS ±0.05% FS ±0.05% FS
Linearity ±1digit ±0.1% FS ±0.05% FS ±0.05% FS ±0.05% FS
Hysteresis max. 0.05% FS
Measurement principle inductive
Temperature drift of the
internal reference sensor
Zero: 0.03% FS / K (0% via zero balance)
Span: 0.03% FS / K
Zero balance • automatic (in settable time spans, possible to swith of), or
• manuel (by pressing the ZERO button)
Long-term stability of the
internal reference sensor
0.1% FS per year (typical)
Temperature ranges Working range +10...+40°C; Storage -10...+70°C
Usable pressure measuring range -10...+110%
Adjustment accuracy 0.01% FS
Adjustment time depending on volume, <5 sec
Measurement media Air, non-aggressive and corrosion-free dry gases
Operation modes • CTRL (controlling pressure)
• MESS (measuring pressure)
• Remote-controlled operation (via interface)
Display TFT touch screen, 7", colour
Interface USB
Supply 24 VDC / 1 A, build-in Li-Ion battery (8h typical)
Pressure ports 6.6 x 11 mm (for flexible hose with 6 mm inner diameter)
Dimension H 191,6 x W 257 x D 271 mm (without handel)
Operating (Menu) languages German, English

The integrated electrical pump can make max. -600 mbar vacuum. If a larger vacuum is required, an external vacuum source is needed, e.g. the pressure test pump  LR-Cal 2941.

Operation Modes

Operation Modes of LR-Cal LCC 200

Measure
The measuring mode is used to measure a gauge or differential pressure. In this operation mode the internal pump of the LR-Cal LCC 200 is deactivated. The applied pressure, measured by the internal reference sensor, is shown on the display.

Control
The control mode is used to calibrate pressure transmitter, pressure switches or pressure gauges. In this operation mode the internal pump of the  LR-Cal LCC 200 is activated, the pressure is adjusted to the set point. The pressure value is shown on the display.

Lecktest
In the Control mode, connected test specimen can be tested to leakage / tightness by pressing the LEAK key.

Accessories

Accessories for LR-Cal LCC 200

Order-Code Description Image
LCC-200-KOFFER

Rigid carrying and storage case with custom foams

Koffer für Druck-Controller LR-Cal LCC 100
LCC-200-KAL-10 Traceable factory certificate of calibration, 10 test points
LCC-200-KAL-DKD DAkkS/DKD certificate of calibration
LCC-200-NT SPARE PART: Mains adapter 110...230 VAC

DOWNLOADs

Datasheet and Manual for LR-Cal LCC 200

Note:
Datasheet and Manual can be downloaded as PDF files. To open PDF files, you need the ADOBE READER. You can get this software free of charge from ADOBE.

Important Note:
This instrument is not available in the moment.

Page Top
Page Bottom

Bitte wählen Sie Ihre Sprache:

LR-Cal Leitenberger Kalibriersysteme
is a trade mark of
DRUCK & TEMPERATUR
Leitenberger GmbH

for calibration equipment
Pressure & Temperature.

190x190-fluid-detail
190x190-gebaeude33
190x190-lsp1200
190x190-lpp700
190x190-gebaeude32
190x190-lsp-tessnow
190x190-lpp1000
190x190-lpc300
190x190-fertigung-1
190x190-lpp08-koffer
190px-ldw-h
190x190-lcc200
190x190-messe1
190x190-fluid200
190x190-lppkit
190x190-pyros650
190x190-lsp1000bm-lpc200
190x190-llc100
190x190-lpc300-finger
190x190-lfc80-atc
190x190-lpp40
190x190-tldmm
190x190-lhm
190x190-lhm-kraft
190x190-lpp60t-tldmm
190x190-lpp60-tldmm
190x190-messe2
190x190-pulsar
190x190-lfc80-atc-detail
190x190-lsp
190x190-llc100anschluss
190x190-fertigung-2
190px-ldw-p
190x190-lcc100
190x190-lfc80
190x190-ldw-hk
190x190-lhm-drehmoment
190x190-lpp60
190x190-lpp60-koffer
190x190-ldm80
190x190-lpp08